CAPINTEC, INC. CAPRAC
®
-t
6-6 WELL COUNTER INITIALIZATION April 2021
TEST SOURCE DATA
A Cs137 test source is used for performing the System Test. The test source data must be
input before a System Test can be carried out.
When shipped from the factory, the CAPRAC
®
-t does not have Cs137 Test Source data
entered.
From Figure 6-4 Advanced Detector Setup Screen, touch the TEST SOURCE button. Figure
6-5 Test Source Setup Screen will appear. If a Test Source has been previously input, the
data for the source will be displayed.
Figure 6-5 Test Source Setup Screen
Note: All fields (S/N:, Calibrated On:, Calibrated Activity:) for the source must be
completed before exiting the Test Source Setup screen or the entered information for
the source will not be saved.
To input (or change) information for the Cs137 Test Source, touch the field’s box and input
the appropriate data for the selected field as described in the following sections.