EasyManuals Logo

Capintec CAPRAC-t User Manual

Default Icon
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #116 background imageLoading...
Page #116 background image
CAPINTEC, INC. CAPRAC
®
-t
April 2021 WELL COUNTER INITIALIZATION 6-35
Trigger Levels
Trigger Levels are count limits that determine when the CAPRAC
®
-t indicates that a
measurement is too high.
Trigger Levels for Background, Wipe, Unrestricted, Sealed Sources and Packages may be
viewed and changed.
The default Trigger Levels set at the factory are listed in Table 6-1 Default Trigger Levels.
Type
Trigger Level
Curies
Becquerels
Background
3,000.00 cpm
50.000 cps
Work Areas
2,000.00 dpm
33.333 Bq
Unrestricted Areas
200.00 dpm
3.333 Bq
Sealed Source
5.000 nCi
185.000 Bq
Packages
200.00 dpm
3.333 Bq
Table 6-1 Default Trigger Levels
The Trigger Levels minimum and maximum values are listed in Table 6-2 Trigger Level
Limits.
Type
Curies
Becquerels
Minimum
Maximum
Minimum
Maximum
Background
60.000 cpm
9,960.0000 cpm
1.0000 cps
166.0000 cps
Work Areas
60.0000 dpm
25,000.0000 dpm
1.0000 Bq
416.6667 Bq
Unrestricted
60.0000 dpm
2,500.0000 dpm
1.0000 Bq
41.6667 Bq
Sealed Source
0.1000 nCi
270.0000 nCi
3.7000 Bq
9,990.0000 Bq
Packages
60.000 dpm
2,500.000 dpm
1.0000 Bq
41.6667 Bq
Table 6-2 Trigger Level Limits
Refer to the appropriate sections below for instructions on editing the Trigger Level for the