EasyManua.ls Logo

Capintec CAPRAC-t - Results

Default Icon
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPRAC
®
-t
10-18 WIPE MEASUREMENT PROCEDURES April 2021
To repeat the measurement, touch the MEASURE button.
If it is desired to change the currently set Nuclide/ROI for the current measurement, touch the
NUCLIDE/ROI button. Figure 10-12 Wipe General Measurement Set Nuclide Screen will re-
appear showing the current settings. Choose a different Nuclide or set the Start and/or End
ROI to the new values as described above in the Set Nuclide/ROI section beginning on page
10-14 and touch the BACK button. Figure 10-14 General Measurement Results Screen will
re-appear showing the newly set ROI highlighted in red, the total and ROI counting rates.
Results
To discard the measurement results, touch the CANCEL button. Figure 10-2 Measurements
Screen will re-appear.
To accept the measurement results, touch the ACCEPT button. Figure 10-15 Wipe General
Measurement Report Screen will appear displaying the results.
Figure 10-15 Wipe General Measurement Report Screen
If a printer is attached to the system, the test results can be printed by touching the PRINT
button.
To exit Figure 10-15 Wipe General Measurement Report Screen,

Table of Contents