CAPINTEC, INC. CAPRAC
®
-t
April 2021 WIPE MEASUREMENT PROCEDURES 10-13
To return to Figure 10-9 Wipe Measurement Results Screen, touch the BACK button.
GENERAL MEASUREMENTS
The General Measurement category is used to measure radiation levels in samples. For
example, this feature can be used to measure liquid samples of blood or urine for lab
procedures. Also, if activities are sufficiently low, radiochromatography strips may be counted
using the General Measurement category.
A General Measurement is not associated with a particular Location. There is no Trigger
Level for General Measurements and no assigned nuclides. Only the counting rate for the
selected channels (energy peaks for any nuclides you may want to look for) is calculated for
General Measurements.
From Figure 10-2 Measurements Screen, touch the GENERAL MEASUREMENT button.
Figure 10-11 Wipe General Measurement Screen will appear.
Figure 10-11 Wipe General Measurement Screen
To exit Figure 10-11 Wipe General Measurement Screen, touch the CANCEL button. Figure
10-2 Measurements Screen will re-appear.