CAPINTEC, INC. CAPRAC
®
-t
April 2021 WIPE MEASUREMENT PROCEDURES 10-7
MEASURE WIPES
Wipes are measured by first selecting them from a list.
From Figure 10-2 Measurements Screen, touch MEASURE WIPES button. Figure 10-6
Measure Wipes Screen will appear.
Figure 10-6 Measure Wipes Screen
If Auto Calibration has not been measured for the current day, the message “Warning
Expired calibration Please run Auto Calibration” will appear. Reference CHAPTER 9:
TESTS, SECTION: AUTO CALIBRATE for Auto Calibration procedures.
If a Background measurement has not been performed for the current day, the message –
“Background Error Background has expired Please run Background for today” will appear.
If the measured Background is higher than the Background Trigger Level, the message
“Background Error Current background exceeds threshold of XXX.X cpm Remove any
sources from well and try again” will appear.
Background can be measured from Figure 10-6 Measure Wipes Screen by touching the
MEASURE BACKGROUND button. (Reference the MEASURE BACKGROUND section
beginning on page 10-3 for Background measurement procedures.) When Background is
measured from this screen, it will be checked against the Background Trigger Level.