CAPINTEC, INC. CAPRAC
®
-t
10-6 WIPE MEASUREMENT PROCEDURES April 2021
Figure 10-5 Background Measurement Results Screen
To see the counts in any channel, use the BLUE ARROW buttons to move the vertical line
cursor to the desired energy channel. DOUBLE ARROW buttons move the cursor
approximately 75 energy channels. SINGLE ARROW buttons move the cursor approximately
7.5 energy channels.
Note: The actual energy channel keV will vary depending upon the Auto Calibration.
When the vertical cursor is moved, the channel energy (keV) and the number of counts in
that energy channel are displayed.
To repeat the background measurement, touch the MEASURE button.
To discard the background measurement results, touch the CANCEL button. Figure 10-2
Measurements Screen will re-appear.
To save the background measurement results, touch the ACCEPT button. Figure 10-2
Measurements Screen will re-appear.