CAPINTEC, INC. CAPRAC
®
-t
April 2021 ACCEPTANCE & QUALITY ASSURANCE TESTS 8-1
CHAPTER 8
ACCEPTANCE & QUALITY
ASSURANCE TESTS
GENERAL ....................................................................................................... 8-1
ACCEPTANCE TESTS ................................................................................... 8-1
QUALITY ASSURANCE TESTS ..................................................................... 8-1
System Test ................................................................................................ 8-1
Contamination Test .................................................................................... 8-2
GENERAL
To insure proper operation of the CAPRAC
®
-t, the following tests should be performed at the
indicated intervals.
ACCEPTANCE TESTS
Perform a Background Measurement as stated in CHAPTER 10: WIPE MEASUREMENT
PROCEDURES, SECTION: MEASURE BACKGROUND.
Verify that the Well Counter data is setup as stated in CHAPTER 6: WELL COUNTER
INITIALIZATION.
Perform Diagnostics as stated in CHAPTER 7: DIAGNOSTICS. If a printer is attached to the
system, verify the printed information is correct.
QUALITY ASSURANCE TESTS
System Test
The System Test should be performed every day. It cannot be performed unless a
Background measurement has been performed for the current day. (Reference CHAPTER 9:
TESTS, SECTION: QUALITY ASSURANCE TESTS, System Test)
If there is a printer connected to the system, the test results can be printed at the end of the
source measurement.