April 2021 WELL COUNTER INITIALIZATION 6-39
Note: Refer to Table 6-2 Trigger Level Limits on page 6-35 for the minimum and
maximum Work Area Trigger Level.
Work Area Counting Time
The default counting time is 60 seconds. If it is desired to use a different Work Area
Counting Time, touch the Count Time: field box.
The numeric keypad will appear. Input the new counting time by touching the
appropriate numbers on the keypad. Touch the ACCEPT button to accept the entered
value. Touch the CANCEL button to abort any changes.
Note: The minimum count time that can be input is 30 seconds. The maximum count
time that can be input is 9,999 seconds.
Work Area Nuclides
Note: Efficiency data must be entered for nuclides to be added. (Refer to the
EFFICIENCIES section on page 6-10)
Up to 10 nuclides may be selected to be looked for in a Wipe sample. The nuclides
selected here will be used as the default nuclides for Work Area Locations.
To add a nuclide to look for, touch a nuclide field box in the Nuclides section of the
screen. The Select Nuclide screen will appear. The screen displays a listing of
nuclides that have an Efficiency value assigned to them (Refer to the
EFFICIENCIES
section on page 6-10) – 10 at a time.
The nuclide list is in alphabetical order. The length of the list will vary depending on
which nuclides have an Efficiency value assigned to them (Refer to the
EFFICIENCIES section on page 6-10). The number of pages (or screens) will be
shown in the upper right corner of the screen indicating the length of the list. To scroll
through the list, touch the DOWN ARROW (▼) button. The next group of 10 nuclides
will be displayed and the UP ARROW (▲) button will appear allowing the user to
scroll up in the list.
To cancel any changes and return to Figure 6-30 Edit Default Work Area Settings
Screen, touch the CANCEL button.
To change the assignment of the chosen Nuclides: field box, do one of the following:
• Touch the Nuclide name on the list on the right. The nuclide will be
highlighted. (If necessary, scroll the list until the desired nuclide is displayed.)
Once a nuclide is selected from the list, an ACCEPT button will appear. Touch
the ACCEPT button to save the change. Figure 6-30 Edit Default Work Area
Settings Screen will re-appear with the selected Nuclide in the chosen
Nuclides: field box.