CAPINTEC, INC. CAPRAC
®
-t
April 2021 TESTS 9-19
To set the Precision Factor, touch the Precision Factor: field box. A numeric keypad
will appear allowing for the entry of a new value.
Input the new value using the keypad and touch the ACCEPT button. Figure 9-15
MDA Test Screen after Background Measurement will re-appear with Precision
Factor: field box populated with the entered value.
To cancel changing the Precision Factor, touch the CANCEL button on the numeric
keypad.
Correction Factor
The Correction Factor used in the calculation must be input. The default Correction
Factor is 0.00.
Note: The allowable range is 0.00-9.00.
To set the Correction Factor, touch the Correction Factor: field box. A numeric
keypad will appear allowing for the entry of a new value.
Input the new value using the keypad and touch the ACCEPT button. Figure 9-15
MDA Test Screen after Background Measurement will re-appear with Correction
Factor: field box populated with the entered value.
To cancel changing the Correction Factor, touch the CANCEL button on the numeric
keypad.