CAPINTEC, INC. CAPRAC
®
-t
April 2021 WELL COUNTER INITIALIZATION 6-9
Note: The recommended calibrated activity is in the 0.1-1.0 µCi range. The calibrated
activity must not be greater than 10.0 µCi (0.37 MBq).
Figure 6-8 Enter Activity Screen
Input the activity value using the keypad and touch the appropriate radio button for the unit of
measure of the Test Source.
Note: The available units will change depending on which unit of measurement is selected
on the Setup screen. i.e. if set for Curies, the available units will be Ci, mCi and µCi; if
set for Becquerels, the available units will be GBq, MBq and kBq.
Touch the ACCEPT button. Figure 6-5 Test Source Setup Screen will re-appear with
Calibrated Activity: field box populated with the entered activity.
To cancel any changes and return to Figure 6-5 Test Source Setup Screen, touch the
CANCEL button.
Once all of the field boxes are filled-in, the screen will be similar to that shown in Figure 6-9
Test Source Setup Screen.