44 Magellan™ 9300i / 9400i Scanners
3. Lower the unit into the counter opening, ensuring that none of the cables
are pinched, pierced or crimped.
4. View the bubble level (if present
1
) located on the scanner's spider assem-
bly to ensure the scanner is level. As shown in the insert of Figure 23, the
air bubble should appear fully within the circle indicated. The air bubble
can touch the circle, but must not overlap it. Apply shims beneath the
scanner flange or adjust the leveling feet to allow level installation.
5. Mount the horizontal surface of the All-Weighs Platter flush with the coun-
tertop to encourage slide scanning rather than lifting.
6. Position the centerline of the scanner read area 20.3 - 25.4 cm (8 - 10
inches) from the edge of the checkstand (cashier side).
7. Reinstall the All Weighs™ Platter and verify that it is flush or just below
flush with the countertop. This is necessary to provide smooth scanning
from either direction. Make adjustments as needed to align the platter with
the counter by moving support rails up or down, or consider installing
screws in positions that will allow their use in adjusting the unit’s position.
Ensure that leveling is maintained if adjustments are made.
You have now completed installation and system integration of the scanner or
scanner-scale. At this point, the scale must be calibrated and certified in accor
-
dance with the requirements of your state and/or local requirements. Refer to
Chapter 6. Calibration Procedures for calibration details.
1. The bubble level is present only on certain models.
NOTE
Failure to install a scanning-scale in a stable and level position will
inhibit weighing, calibration and zeroing operations. The platter
MUST make unobstructed contact with all of its supports for proper
weighing operation.