6. Operation AUSTROMAT
®
664 iSiC
®
/ 674 baSiC
2
110 Operating Instructions (Version 03/2018)
6.11. After a power failure
NOTE:
If a running program is interrupted by a temporary power failure (less than 2 hours), the
firing furnace attempts to properly terminate the program when the power comes back
on. The quality of the firing results is not guaranteed in this case. Make sure to take
particular care to inspect the firing results.
After a power failure, the control remembers the status of the firing furnace at the time of the
interruption for up to two hours.
If the power comes back on within two hours, the firing furnace returns to the status at the
time of the power failure. A firing program is continued if the temperature drop in the firing
chamber does not exceed 150°C.
Otherwise, the AUSTROMAT
®
664 / 674 asks you if you want to continue the program.