4. Start-up AUSTROMAT
®
664 iSiC
®
/ 674 baSiC
2
36 Operating Instructions (Version 03/2018)
5. Under "Gateway", enter the gateway address that was assigned to the firing furnace by the
network administrator.
6. Deactivate the DHCP function.
7. Click on [OK].
The firing furnace now has a fixed IP address and can be accessed using a browser or the VNC
program at this address.
NOTE:
The new IP address is only activated after the firing furnace is restarted.
Now set up the authorizations for accessing the firing furnace (see section 4.5.5).
4.5.5. Specify authorizations
After connecting the firing furnace to a network or individual computer, each network participant
or computer user can access the firing furnace control. You can restrict the access the firing
furnace to authorized persons by providing a user ID consisting of a user name and password.
1. In the "Setup" menu, open the "Login data" function (see section 5.8.4.4).
2.
Define the p
assword: In the "Password" field, enter the password,
3. Activate the functions "Start/stop functionality" and "Function keys via the Web server".
When opening the firing furnace control through a browser or VNC program, the password has
to be entered to obtain access to the firing furnace, i.e., the firing furnace is password-protected.
Communicate the password to the persons who are authorized to access the firing furnace.
NOTE:
If the password becomes lost, contact your service partner or directly contact DEKEMA
(see page 2).
4.5.6. Access through a browser or a VNC program
You can now operate the firing furnace using either a browser or the VNC program.