Step 3 "T1200": Wait time 1200 s.
Step 4 "T010.C1475": Set firing temperature (1475 °C) with a rate of
10°C/min (900 °C to 1475 °C)
Step 5 "T4500": Wait time 4500 s.
Step 6 "T010.A900": Set object temperature (900 °C) with a rate of
10°C/min (1475°C to 900°C) The lift travels downward somewhat
while the firing chamber is being heated.
Step 7 "T030.A250": Set object temperature (250 °C) with a rate of
10°C/min (900 °C to 275 °C) The lift travels downward somewhat
while the firing chamber is being heated.
Step 8 "C0": Turning off the heater
Step 9 "L0": The lift moves down all the way.
Step 10 "T2": The acoustic signal for the end of the program sounds. The
firing number is automatically increased by one.
Classic programs
The individual program steps are displayed in the "CVTL" syntax and processed from left to
right in the given sequence.