EasyManuals Logo

Dräger Infinity Delta User Manual

Dräger Infinity Delta
614 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #302 background imageLoading...
Page #302 background image
17-12 DELTA/DELTA XL/KAPPA VF8
17 P
ULSE OXIMETRY (SPO2)
Monitor with Nellcor Logo - Status Messages
Message Possible Cause Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
falls outside the
current upper or lower alarm limits.
Check the patient and treat, if
necessary.
PLS > #
PLS < #
The patient’s pulse rate falls outside
the current upper or lower alarm
limits.
SpO
2
< Low limit
value - 20%
The patient’s SpO
2
falls below the
current lower alarm limit by 20% or
more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
Unrecognized
Sensor
The monitor does not recognize the
connected sensor as valid.
Check for defective or unapproved
sensor.
Replace the sensor.
Disconnect/reconnect the pod and
replace with a good sensor.
Contact the hospital’s technical
personnel or DrägerService.
SpO
2
Artifact/
Sensor/Site?
A persistent artifact
Wrong sensor and/or site used
High pulse amplitude
Sensor not properly placed
Electrical/optical interference
Light is blocked
Make sure the SpO
2
sensor is the
correct type and is properly
attached to the patient.
Make sure the patient is not
moving, and all cables are properly
connected.
Make sure no electrical or optical
interference is present.
Check the sensor for blocked light.
(May be due to nail polish)
Change the site/location
periodically. (for example finger vs.
forehead, etc.)
Contact DrägerService.
SpO
2
Weak signal/
Sensor Off?
The pulse amplitude is too low.
Weak signal/weak pulse.
Physiological:
- Poor perfusion
- Low body temperature
- Sensor off the patient
Check the patient and treat, if
necessary.
Make sure the SpO
2
sensor is
properly attached to the patient and
all cables are properly connected.
Relocate the sensor to another
extremity.
Contact DrägerService.

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms

Related product manuals