PATIENT P REPARATION
VF8 DELTA/DELTA XL/KAPPA 17-7
Patient Preparation
The accuracy of SpO
2
monitoring depends largely on the strength and quality of the
SpO
2
signal.
If a finger is used as a monitoring site, remove any nail polish. Trim the patient's
finger nails, if necessary, for better sensor placement. Use only Dräger provided
sensors and apply them per the sensor manufacturers recommendation (see page C-13).
Ambient light can interfere with pulse oximetry measurements if the sensor is not
properly attached, causing erratic measurement or missing values. Observe proper
sensor placement and cover the sensor with opaque material if interference due to
ambient light is suspected.
1. Select the sensor type and size best suited for your patient.
2. If the sensor is reusable, clean it before and after each patient use.
3. Position the sensor correctly and attach it to your patient.
4. Connect the sensor to the patient cable. For the appropriate patient cable for
your specific SpO
2
device, see page C-15.
5. Inspect the sensor application site frequently. A sensor that is too tight may
damage the tissue by impeding blood flow or overheating the skin. Do not use
a damaged sensor.
NOTE: Read the instructions provided with the sensor for optimal application
techniques and for safety information. Never use damaged sensors.