27-6 DELTA/DELTA XL/KAPPA VF8
27 B
ODY TEMPERATURE
Temperature Setup
To access the temperature setup menu
Click on the desired temperature parameter box (if displayed);
or
1. Press the
Menu fixed key.
2. Click on
Patient Setup.
3. Click on
Parameters.
4. Click on TEMP, TEMP2 or TEMP3, depending on which temperature signal
you wish to monitor (see page
27-4).
The TEMP setup menu displays only two items:
TEMP display – Configures the bottom half of the parameter box to display
either the reading of the second temperature probe (b) or the difference
between the first probe’s reading and the second (DT, the delta value).
TEMP alarms – Accesses temperature alarm settings on the Alarm Limits table
(see page 5-6).
Status Messages
Message Possible Cause Suggested Action
> #
< #
The temperature exceeds the
upper or lower alarm limits.
Check the patient and treat, if
necessary.
Check equipment and replace if
necessary.
Out of Range (High)
Out of Range (Low)
The temperature value is greater
or less than the measuring range.
Can’t Derive
T(#) The cable is defective or
unplugged.
Check equipment and replace if
necessary.
Connect the second temperature
probe.
Unplugged Check the equipment and replace if
necessary.
H/W Failure Temperature circuitry failure. Contact DrägerService.