10-4 DELTA/DELTA XL/KAPPA VF8
10 ST M
ONITORING
ST Display
When ST monitoring and ECG monitoring are enabled, the ST
parameter box appears just below the HR parameter box on the main
screen.
You can enable and disable ST monitoring on the ST Analysis menu
or ECG setup menu (page 8-18). ST deviation values are displayed in
the same format as strip recordings, where 1 millimeter (mm) on the
grid corresponds to 0.1 millivolt (mV).
ST Analysis Setup
The ST analysis menu allows you to execute most of the functions involved in
analyzing the ST segment.
To open the ST Analysis menu
Click on the ST parameter box (if displayed),
or
1. Press the
Menu fixed key on the front of the monitor.
2. Click on
Patient Setup.
3. Click on
Parameters. A list of available parameters appears.
4. Click on
ST.
NOTE: ST does not appear in the parameter list if ST monitoring has been disabled.