29-6 DELTA/DELTA XL/KAPPA VF8
29 C
LEANING AND DISINFECTING
To clean the MPod – Quad Hemo
1. Disconnect the MPod – Quad Hemo from the bedside monitor.
2. Clean the MPod – Quad Hemo with a gauze pad moistened with soapy water
or with an approved cleaning agent.
3. Dry thoroughly with a lint-free cloth.
To disinfect the MPod – Quad Hemo
1. Disconnect the MPod – Quad Hemo from the bedside monitor.
2. Disinfect the MPod – Quad Hemo with a gauze pad moistened with diluted
alcohol.
3. Dry thoroughly with a lint-free cloth.
SpO
2
Clean reusable SpO
2
sensors by wiping them with a gauze pad moistened with a soap
solution. To disinfect sensors, wipe using a cloth moistened with a 70% alcohol
solution. Dry thoroughly with a lint-free cloth before applying to patient.
Reusable SpO
2
Sensor
Refer to the cleaning instructions and recommendations provided with the sensor.
Infinity Masimo SET SpO
2
Pod
1. To clean the Masimo SET SpO
2
pod disconnect it from the monitor.
2. Clean the pod with a gauze moistened in enzymatic detergent or a solution of
green tinctured soap and water. Dry thoroughly with a lint-free cloth.
CAUTION:
Do not use organic solvents.
Do not sterilize by steam, heat, radiation or ETO.
Do not use sharp objects.
Make sure that NO liquid enters the pod.