EasyManua.ls Logo

Dräger Infinity Delta - Quick Reference Table - bis Setup Menu

Dräger Infinity Delta
614 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25-6 DELTA/DELTA XL/KAPPA VF8
25 B
ISPECTRAL INDEX (BISX) MONITORING
Quick Reference Table – BIS Setup menu
Menu Item Description Available Settings
Show all
parameters
Opens a separate menu with all BISx
parameters displayed.
BIS: Bispectral Index
(range = 0 - 100)
continuous processed EEG
parameter that correlates to
a patient’s level of
consciousness, where
100=awake and 0=flatline
EEG.
SQI: Signal Quality Index
(Range = 0 - 100 %)
a measure of the signal
quality for the EEG channel
source, which is calculated
based on impedance data,
artifact, and other variables.
It is displayed in the form of
a simple bar graph, located
at the bottom of the
Parameter box, above EMG.
EMG: Electromyograph Indicator
(Range = 0 - 100 dB)
The power (in decibels) in
the frequency range 70-
110Hz is labeled “EMG”.
This frequency range
contains power from muscle
activity as well as power
from other high-frequency
artifacts. It is displayed as a
simple bar graph, located at
the bottom of the parameter
box, below SQI.
BSR: Suppression ratio
(range = 0 - 100 %)
A calculated parameter to
give the user an indication
when an isoelectric (flatline)
condition may exist.
Suppression ratio is the
percentage of time over the
last 63-second period that
the signal is considered to
be in a suppressed state.
For example BSR=11
(isoelectric over 11% of the
last 63-second review, or 7
seconds).
SEF: Spectral Edge Frequency
(range = 0.5 - 30.00 Hz)
The frequency at which 95%
of the total power lies below
it and 5 % lies above it.

Table of Contents

Other manuals for Dräger Infinity Delta