ELECTROMAGNETIC COMPATIBILITY (EMC)
VF8 DELTA/DELTA XL/KAPPA B-7
Recommended separation distances
Recommended separation distances between portable and mobile RF communications
equipment and the equipment
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter in meters
150 kHz - 80 MHz 80 MHz to 800MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
NOTE:
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
d1.2
P= d1.2
P=
d 2.3
P=