DRÄGER INFANT INCUBATOR CALEO
VF8 DELTA/DELTA XL/KAPPA 28-27
C2000/C2000e Status Messages
Dräger Infant Incubator Caleo
The monitor, when connected to a Caleo infant incubator via MIB, displays skin
temperature, air temperature, humidity, oygen levels and weight.
There are two parameter boxes associated with the Incubator, each with the ability to
display up to three parameters. The Incub1 parameter box appears by default when the
device is connected. However, the user can select Incub1or Incub2 separately from the
parameter pick list depending on the number of parameter boxes selected in the menu
selection.
Message Condition Suggested Action
Duplicate Device
Connected
Multiple devices are connected. Make sure only one incubator is
connected to the monitor.
Tair Out of Range The Tair value is out of range. Check the patient and treat, if necessary.
Check the connections.
Unplug and reconnect the incubator.
Power-cycle the monitor or undock and
redock the monitor.
Call the manufacturer.
T1skin Out of
Range
The T1skin value is out of
range.
T2skin Out of
Range
The T2skin value is out of
range.
Wt Out of Range The weight value is out of
range.
R.H Out of Range The R.H value is out of range
(low).
%O2 Out of Range The %O2 value is out of range.
NOTE:
The Caleo infant incubator is only available in neonatal patient category.
T1 skin and T2 skin are plotted on the same graphical trend channel.
Label Parameters Units
Tair Air temperature °C (°F)
T1skin &
T2skin
Skin temperature °C (°F)
Wt Weight g (oz)
R.H Humidity %
%O2 Oxygen %

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms