23-24 DELTA/DELTA XL/KAPPA VF8
23 S
CIO® FOUR MODULES
Dual Agent Display (Scio Four Oxi plus & Scio Four plus
only)
If single agent display is selected, the inspired and expired concentrations for two
agents is displayed. The MAC value is the total expired MAC value for the two agents
and N
2
O.
If the combined O
2
/N
2
O/Agent display is selected, the primary agent is displayed.
Maintenance and Repair
The Scio module requires routine cleaning (see chapter 29). Monitoring accessories
such as sampling tubing, T-connectors, water traps, and fan filters are not reusable and
need to be replaced at regular intervals.
General equipment inspection and maintenance is required. Once a year, check all
devices, accessories, and cables for damage and test the ground resistance, chassis and
patient leakage currents as well as all alarm functions. Make sure that all safety labels
are legible. Maintain a record of these safety inspections.
Changing the Sampling Tubing and T-Connector
The sampling tubing and T-connector that connect the Scio module to the patient’s
airway are not reusable. They must be replaced under the following conditions:
A new patient is connected to the module
The cleanliness of the tubing or connector is suspect or compromised
NOTE: Dual agent is available when the Scio module is connected via a Scio
connecting cable (PN: 7876878 or 7876886) to the X3 or Scio/ISD connector, or via a
Scio Direct connect cable (PN: 6871581) to the X8 connector.
NOTE: When two agents are detected, the one with the higher expired MAC value is
the primary.
NOTE: Use only sidestream sampling lines provided by Dräger, otherwise the correct
function of the device may be compromised.