SCIO S ETUP
VF8 DELTA/DELTA XL/KAPPA 23-23
Quick Reference Table – Combined Display Setup
Click on the following items to execute setup functions for combined (O
2
/Agent/N
2
O)
MultiGas monitoring and display.
The Combined (O
2
/Agent/N
2
O) Setup Menu
Menu Item Description Settings
Waveform Selects a waveform for display. O
2,
Agent
MultiGas
Parameter
Enables the combined MultiGas parameter
box display (see page 23-22).
•ON
•OFF
O2 Scale Sets O
2
waveform scale (see page 23-13). 50 %, 100 %
N2O Display Displays N
2
O values in the O
2
/N
2
O
parameter box (see page 23-16).
•ON
•OFF
Agent Display Displays a separate agent parameter box
(see page 23-21).
•ON
OFF
NOTE: During the combined display, the N
2
O and agent displays are automatically turned on and
their selections are ghosted on the combined menu. You can only access these selections and turn
the displays off, once you have turned off the combined MultiGas parameter display.
Agent Scale Sets the agent waveform scale. 1-20 % in increments of 1
MultiGas Zero Manually zeroes the Scio module.
NOTE: During zeroing, the monitor
temporarily blanks Scio parameter values.
Not applicable
Auto Zero Delay Delays automatic zeroing for 5 minutes (see
page 23-13 for detailed information).
Not applicable
O2 Calibration Selects desired O2 calibration
*See page 23-12 for further information on
calibration functions.
1 Point Cal.*
2 Point Cal.*
Last O2 Cal.*
O2 Alarms Accesses O
2
alarms in the Alarm Limits table (see chapter 5)