2-14 DELTA/DELTA XL/KAPPA VF8
2 M
ONITOR SETUP
Unit Manager
The Unit Manager menu lets supervisory personnel configure monitoring functions
for the clinical staff. Access to this menu is restricted by a password. To open the Unit
Manager menu:
1. Press the
Menu fixed key to open the Main menu.
2. Click on
Monitor Setup.
3. Click on Unit Manager. A data entry box appears.
4. Scroll through the numbers and click successively on the single digits of the
clinical password. If you make a mistake, click on
Backspace and try again.
5. Click on
Accept to open the Unit Manager menu. Available functions are
described in the table that follows.
Pulse Tone
Volume
Sets the pulse tone volume
(see page 2-8).
10% - 100%
The Unit Manager Menu
Menu Item Description Available Settings
The Alarm Control Submenu
This menu allows the unit manager to configure alarm annunciation. Open the Unit Manager menu,
click on Alarm Control, then follow the procedures outlined in this table to execute the indicated
functions.
All Alarms OFF
Key
Determines whether clinical personnel
can suspend alarms by using the All
Alarms Off fixed key on the front of the
monitor.
ON (default) – Pressing All
Alarms Off fixed key suspends
alarms.
OFF – Pressing All Alarms Off
fixed key triggers an error tone.
All Alarms OFF
Time
Sets the alarm suspension time.
WARNING: If No Timeout is assigned
to the ‘alarm off period’, no counter
appears and alarms remain
deactivated until you turn them on
again.
1, 2 (default), 3, 4, or 5 min – A
timer at the top of the screen
indicates the amount of time
remaining in All Alarms OFF
Time.
No Timeout – Alarms are
suspended indefinitely; no timer
appears.