CALIBRATION
VF8 DELTA/DELTA XL/KAPPA 20-7
Quick Reference Table – etCO2 Setup
The following table explains Microstream pod etCO
2
setup functions.
Calibration
Calibration of the Microstream pod is not necessary unless the calibration check fails.
A calibration check should be performed when the monitor displays the advisory
message Calibration Required. An initial calibration is requested after the first 1,200
operating hours. Afterwards, calibration is required annually or at 4,000 operating
hours, whichever is first.
Calibration is a Biomed function and is password protected. Consult your hospital’s
technical personnel for further information.
Function Description Settings
Scale Determines size of currently displayed
waveform.
• 40 mmHg
• 60 mmHg
• 80 mmHg
• 100 mmHg
Respiratory
Sweep Speed
Sets waveform sweep speed on screen
display.
• 6.25 mm/s
• 12.5 mm/s
• 25 mm/s
• 50 mm/s
Averaging Sets interval for CO
2
measurements.
NOTE: The monitor reports the
maximum value of etCO
2
during the
specified sampling interval.
• Breath-Breath
• 10 s
• 20 s (default)
• 30s
etCO2 Alarm Accesses Alarm Limits table, where you
can set upper and lower alarm
thresholds.See chapter 5, Alarms, for
information about setting and displaying
alarm limits.
Not applicable
BTPS (body
temperature,
pressure,
saturation)
User selectable correction factor. • On (default)
•Off
Last Cal Displays the last calibration date of the
pod (read only).
Not applicable