RESP PRECAUTIONS
VF8 DELTA/DELTA XL/KAPPA 12-3
RESP Precautions
For general safety precautions regarding electrosurgery, see “Safety Considerations
on page 6 of these Instructions for Use.
WARNING:
The safety and effectiveness of the respiration
measurement method in the detection of apnea,
particularly the apnea of prematurity and apnea of
infancy, has not been established.
This device does not monitor obstructive apnea.
Patients at risk for respiratory crises should be
observed closely.
The monitor reports an apneic event when no breaths
are reported within the established apnea time period.
Therefore, do not rely on impedance respiration
monitoring as the sole method for detecting cessation
of breathing. Dräger recommends the monitoring of
additional parameters that indicate the patient’s
oxygenation status, such as etCO
2
and SpO
2
. Heart rate
limit alarms should also be enabled and set
appropriately.
While the ESU block or the MultiMed Plus OR cable is in
use, impedance respiration monitoring is inoperative
and the detection of pacemaker spikes could be
compromised. If pacemaker detection is enabled, the
ESU interference may be detected as pacemaker spikes.
Large amplitude pacemaker pulses (100 mV or greater)
may interfere with the monitors ability to measure or
detect respiration.

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms

Related product manuals