12-10 DELTA/DELTA XL/KAPPA VF8
12 R
ESPIRATION
To display the OCRG screen
1. Set the patient type to Neonatal (see page 2-4).
2. Connect SpO
2
sensors, HR leads, and respiration or etCO
2
leads.
3. Set the apnea time on the RESP menu (see page 12-7).
4. Press the
Fast Access fixed key.
5. Click on
OxyCRG to display the OCRG screen.
Scale
To change the bbHR scale
1. Using the rotary knob, highlight the value at the top left corner of the grid and
click.
2. Dial to desired scale setting and click.
Values are shown in the following table (you can modify only the HR scale).
Cursor
When you click on Cursor, a vertical bar appears on the trend area of the screen, and
cursor time is displayed in the upper right corner. The numbers on the left side of the
screen no longer represent scale values, but rather parameter values at the time marked
by the cursor. The monitor continues to display current (real-time) values on the right
of the screen. When you move the cursor to the right or the left with the rotary knob,
the cursor-time values and the cursor time are modified and displayed accordingly.
Parameter Scale Definition
bbHR 10-180 bpm (lowest setting)
130 - 300 bpm (highest setting)
Highest (max) and lowest (min) bbHR value over
the last three minutes
SpO
2
50 -100 % Lowest saturation value over last three minutes
tpO
2
/
tpO
2
*
30 - 110 mmHg
(4 - 16 kPa)
Lowest tpO
2
/tpO
2
* saturation value over last three
minutes