OXYCRG (OCRG) MONITORING
VF8 DELTA/DELTA XL/KAPPA 12-11
Review Summary Screen Overview
Use the Review Summary screen to view stored bradycardia, SpO
2
, and apnea events.
This screen also allows you to access data associated with each event. The monitor
stores up to 75 total events, but only retains associated data from the 50 most recent
OCRG events (new data overwrites the old).
When an OCRG trigger event occurs, an orange bar appears in the appropriate
parameter graph on the OCRG Review Summary screen. Whenever an OCRG event
occurs, the monitor automatically captures bradycardia, SpO
2
, and apnea event data
(if available). Event data associated with the trigger event appears in the other two
rows (BRDY, SpO
2
, or Apnea) color coded as follows:
Bradycardia - green
SpO
2
- white
Apnea - blue
Accessing Review Summary Screen
To access the OCRG review screen
Enter the alarm history table and click on quick access key OxyCRG Review.
or
Click on the Summary button on the OCRG screen.
.
NOTE: The OCRG Review Summary screen does not time-out and does not update
automatically. This screen remains displayed until the user removes it.To update the
display, click the back arrow, then click the Summary button again.
1
Cursor
2
Button for scrolling to
the right page by page
3
Cursor button
4
Horizontal scroll bar
5
Time interval setting
6
Button for scrolling to
the left page by page
7
Time bar
8
Back arrow