17-6 DELTA/DELTA XL/KAPPA VF8
17 P
ULSE OXIMETRY (SPO2)
Monitor Without Masimo or Nellcor Logo - Hardware Setup
1
Monitor without Masimo or
Nellcor logo (see
compatibility in note below)
2
MultiMed connector
3
Infinity Kappa
4
SpO2 connector on
MultiMed
5
Masimo Procal+
Intermediate LNOP cable
6
Masimo LNOP SpO2 sensor
7
Dräger/Nellcor SpO2 sensor
8
Dräger/Nellcor Intermediate
cable
9
SpO2 connector on
MultiMed Plus cable
10
Infinity Delta/Delta XL
WARNING: Only use Dräger SpO
2
intermediate cables and
SpO
2
sensors. Do not use other manufacturer’s cables or
sensors.
NOTE: A monitor that does not have a Masimo SET or Nellcor Oximax logo in the
area near the rotary knob is compatible with:
OxiSure via SpO
2
connector on various MultiMed cables (see above graphic).
Masimo SET pod via Masimo intermediate cable (see page 17-4).
Nellcor Oximax pod via Nellcor intermediate cable (see page 17-5).