29-4 DELTA/DELTA XL/KAPPA VF8
29 C
LEANING AND DISINFECTING
NBP
Wipe the NBP cuff with a cloth moistened with soap and water or a solution based on
household bleach (1:10), alcohol, or phenol.
IBP
Transducers
Always handle transducers and other pressure accessories with great care. Do not
apply excessive pressure to a transducer diaphragm. Do not subject transducers to
water, steam, hot air sterilization, ether, chloroform, or similar chemicals. Always
protect the connector from moisture.
Consult the documentation supplied with your transducer for specific cleaning and
sterilizing instructions.
Hemodynamic Pod Transducer Plate
Remove the transducer mounting plate from the front of the hemodynamic pod. Wash
the plate with hot soapy water.
CAUTION: The NBP cuff can be immersed in cleaning solution, but do not allow the
solution to enter the NBP hose. The warranty is void if cleaning solution is allowed to
enter the hose or the cuff.

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms