MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-33
Monitoring Specifications
ECG
WARNING: The following parameters are not monitored in
neonatal mode: arrhythmia, sidestream etCO2, cardiac
output, ST, and all MultiGas parameters.
Display: Up to 12 leads
Available leads:
Adult and pediatric (regular) with TruST
off:
I, II, III, aVR, aVL, aVF, V, V+, V1-V6
(aVR/aVL/aVF/V only with 5-lead set; V+ only with 6-
lead set; all leads only with 12-lead set)
Adult and pediatric with TruST on: I, II, III, aVR, aVL, aVF, V2, V5, dV1, dV3, dV4, dV6
NOTE: TruST leads are indicated by a prefix “d” before
the V lead.
Neonatal: I, II, III, aVR, aVL, aVF, V, V+
(aVR/aVL/aVF/V only with 5-lead set; V+ only with 6-
lead set)
Measurement range: 15 to 300 1 b/min
Accuracy: ±2 bpm or ±1 % (whichever is greater)
Display update period: 2s.
QRS detection: Amplitude:0.5 to 5.0 mV
Duration:70 to 120 ms (adult and pediatric)
40 to 120 ms (neonatal)
Frequency ranges: filter = Monitor:0.5 to 40 Hz
filter = ESU:0.5 to 16 Hz
filter = OFF:0.05 to 40 Hz
NOTE: Printed ST and Rest-ECG reports conform to
EC-11 diagnostic bandwidth requirements.
Degree of protection against electric
shock:
Type CF
Defibrillation protection: In accordance per IEC 60601-2-27
Arrhythmia detection: Adult and Pediatric:Yes
Neonatal:No
Pacer detection: Adult and pediatric:Yes, on leads I, II, or III
Neonatal:No
Detects pacemaker pulses with the following characteristics:
Amplitude ±2 to ±700 mV
Width (dp) 0.2 to 2.0 ms
Rise/Fall times (min.) 0.1 d
p
, 100 ms
Overshoot (min.) 0.025 a
p
, 2 mV
Recharge time constant 4 to 100 ms