B-28 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
Trident (NMT) Pod
BISx Pod
Physical Attributes
Size (H x W x T) 36 x 132 x 140 mm (1.42 x 5.19 x 5.51 in)
Weight 595 g (21 oz.)
Connections PodCom: for connection to monitor
•NMT Cable
EEG: for connection to an EEG pod or BISx pod.
Electric shock
protection:
Type CF
Mode of operation:
Continuous
Standards: IEC 60601-1
IEC 60601-2-40: Particular Requirements for Safety - Specification for
Electromyographs and Evoked Response Equipment.
Defibrillator proof?
No
Environmental Requirements
Temperature range:
Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 40 °C (–4 to +104 °F)
Relative humidity:
Operating:20 to 90 %, non-condensing
Storage: 20 to 90 %, (with packaging)
Atmospheric pressure
Operating:525 to 795.1 mmHg (70 to 106 kPa)
Storage: 375 to 795.1 mmHg (50 to 106 kPa)
Physical Attributes
Size (H x W x T) 114.3 x 95.3 x 63.5 mm (4.5 x 3.75 x 2.5 in)
Weight (without
PodCom cable)
227 g (8 oz.)
Connections PodCom: for connection to monitor
Power source:
Powered directly from monitor
Protection against
electric shock:
Type BF
Mode of operation:
Continuous

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms

Related product manuals