EasyManua.ls Logo

Dräger Infinity Delta - Monitoring Accessories

Dräger Infinity Delta
614 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-21
Monitoring Accessories
etCO
2
Module
Physical Attributes
Size (H x W x D):
Module: 150 x 93 x 65 mm (5.9 x 3.6 x 2.6 in)
Capnostat III sensor:
33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in)
Weight:
Module: 0.5 kg (1.1 lb)
Capnostat
TM
III sensor:18 g
Connections:
Sensor connector, female luer side-stream sampling port, male luer
sample exhaust port
Adult airway adapter dead
space:
<5 cc
Neonatal airway adapter
dead space:
< 0.5 cc
Moisture resistance:
Airway adapter can be immersed in water without damage
NOTE: Capnostat
TM
III sensor size and weight exclude cable.
Electrical Specifications
Power source:
Powered directly from monitor
Protection against electric
shock:
Type CF
Mode of operation:
Continuous
Environmental Requirements
Temperature range:
Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 50 °C (–4 to 122 °F)
Relative humidity:
Operating:20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure:
Operating:525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)