B-22 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
etCO
2
Pod
Physical Attributes
Size (H x W x D):
Pod: approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Capnostat
TM
III Sensor: 33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in)
Weight:
Pod: 0.5 kg (1.1 lb)
Capnostat
TM
III sensor: 18 g
Connections:
Sensor connector, female luer side-stream sampling port, male
luer sample exhaust port
Adult airway adapter dead
space:
< 5 cc
Neonatal airway adapter dead
space:
< 0.5 cc
Moisture resistance:
Airway adapter can be immersed in water without damage
Biotoxicity:
None per ASTM F720 (MTL #2009)
NOTE: Capnostat
TM
III sensor size and weight exclude cable.
Electrical Specifications
Power source:
Powered directly from monitor
Protection against electric
shock:
Type CF
Mode of operation:
Continuous
Defibrillation protection:
Defibrillation-proof applied part per IEC 60601-1
Environmental Requirements
Temperature range:
Operating: 10 to 40 °C (50 to 104 °F)
Storage: –20 to 50 °C (–4 to 122 °F)
Relative humidity:
Operating: 20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atm. pressure:
Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
NOTE: Reading complies to BTPS = body temperature 37 °C, ambient barometric pressure of
750 mmHg, and relative humidity of 100 %.
Regulatory Requirements
Standards: IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable
sections only, Oxygen analyzers for monitoring patient breathing
mixtures); EN ISO 21647:2009