MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-23
etCO
2
/Respiratory Mechanics Pod
Physical Attributes
Size (H x W x D):
Approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Weight:
Pod: 0.54 kg (1.2 lb)
Flow sensor: 59 g
Degree of protection against
ingress of water:
Pod is resistant to spills only.
NOTE: For information on the Capnostat
TM
III sensor, see B-22. For more information about airway
adapters used with the etCO2/respiratory mechanics pod, see B-50.
Electrical Specifications
Power source:
Powered directly from monitor
Protection against
electric shock:
Type CF
Mode of operation:
Continuous
Defibrillation protection:
Defibrillation-proof applied part per IEC 60601-1
Environmental Requirements
Temperature range:
Operating: 10 to 40°C (50 to 104 °F)
Storage: –20 to 50 °C (–4 to 122 °F)
Relative humidity:
Operating: 20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atm. pressure:
Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
NOTE: Reading complies to BTPS = Body temperature 37 °C, ambient barometric pressure of
750 mmHg, and relative humidity of 100 %.
Regulatory Requirements
Standards: IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable
sections only, Oxygen analyzers for monitoring patient breathing
mixtures); EN ISO 21647:2009