SpO2 Desat Limit This feature disables the SpO
2
desaturation alarm or sets a threshold at
which point a desaturation alarm of high
priority is triggered. For example, if the
lower alarm limit for SpO
2
is set at 95%
and the desaturation limit is set at –5, the
desaturation alarm will be triggered
when the patient’s SpO
2
value goes
below 90%.
NOTE: This selection is available for all
patient categories and is protected by a
clinical password.
–5, –10 (default), –15,
–20
•OFF
SpO2 Alarm Accesses SpO
2
alarms on the Alarm
Limits table (see page 5-6).
Not applicable
Message Possible Cause Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
falls outside
the current upper or lower alarm
limits.
Observe the patient and treat, if
necessary.
PLS > #
PLS < #
The patient’s pulse rate falls
outside the current upper or
lower alarm limits.
SpO2 < Low limit
value - 20%
The patient’s SpO
2
falls below
the current lower alarm limit by
20% or more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
Transparent Too much light is reaching the
sensor’s light detector because
because the sensor is either
disconnected or too much
ambient light is present.
Make sure the sensor is properly
attached to the patient’s finger.
Remove the light source.
Cover the sensor with opaque material.
Make sure no ambient light can reach
the detector.
Contact DrägerService.