ETCO2 DISPLAY FEATURES
VF8 DELTA/DELTA XL/KAPPA 20-5
Capnograms
The monitor also displays an instantaneous CO
2
reading as a waveform or capnogram.
A typical capnogram is shown below:
You can use capnograms to troubleshoot problems with equipment or monitor
configuration as well as to monitor a patients clinical status. The following table
shows some of the more common problems identifiable through capnogram analysis:
Capnogram Description Immediate and possible causes
Alveolar plateau shows
downward slope that
merges with descending
limb
Inadequate seal around endotracheal tube
Leaky or deflated endotracheal or
tracheostomy cuff
Artificial airway that is too small for the
patient
Elevated waveform
baseline with
corresponding increase in
etCO
2
level
Rebreathing
Insufficient expiratory time
Faulty expiratory valve
Inadequate inspiratory flow
Malfunction of a CO
2
absorber system
Partial rebreathing circuits
1
Expiratory plateau (level of CO
2
in lungs ceases to increase significantly)
2
End-tidal concentration point (end of expiration phase, where etCO
2
is
measured)
3
Onset of inspiration phase
4
Expiration phase
5
Baseline during inspiration