DATA ARCHIVE APPLICATIONS
VF8 DELTA/DELTA XL/KAPPA 1-27
Event Recall
The monitor stores monitoring data (waveforms and parameter values), alarm
conditions, and arrhythmia events on the Event Recall screen. You can view up to 50
stored events, each containing 20 seconds of data, with associated date and time
stamps. Events are stored on a first-in, first-out basis. When event storage is full, the
monitor deletes the oldest events to make room for new ones. All stored events are
deleted whenever you discharge a patient, reset the monitor, or temporarily lose
power.
To access the Event Recall screen
1. Press the Fast Access fixed key.
2. Click on
Event Recall to display the Event Recall screen.
1
Time of capture
6
View1:
2
Parameter values at time of capture
7
Parameter display2:
3
Print report
8
Parameter labels
4
Requests recording
9
Waveform delay and speed
5
Saves/Deletes events
1
View: All - displays all stored events; Manual - displays manually stored events;
Alarm - displays alarm events only; BRDY - displays bradycardia events only;
Desat (Neonatal only) - displays desaturation events only
2
Parameter display: Prev - displays previous set of (2) parameters; Next - displays
next set of (2) parameters