EasyManua.ls Logo

Dräger Infinity Delta - Status Messages

Dräger Infinity Delta
614 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 10-11
Status Messages
Message Possible Cause Suggested Action
ST <x> Out of
Range High
ST <x> Out of
Range Low
The ST algorithm has calculated
values ±15 mm (or ±1.5 mV)
outside the high or low end of the
ST measurement range.
Check the isoelectric and ST measuring
points.
Observe the patient and treat if clinically
indicated.
Cannot Analyze
ST
The monitor cannot determine
ST values because of the
following conditions:
Absence of normal beats
•Artifact
Perform a relearn.
Calm the patient.
Check electrodes; re-apply, if necessary.
Make sure that the patient’s skin is properly
prepped.
Isolate the patient from auxiliary equipment
if possible.
ST <x> > <#>
ST <x> < <#>
The ST value is outside the
upper or lower alarm limit.
Observe the patient carefully and treat if
clinically indicated.
Change the alarm limits.
ST <x> Lead
Invalid
Bad electrode contact or faulty
lead wire.
Inspect and replace defective cables and
wires.
Reapply gel on reusable electrode(s).
Reapply electrodes. Make sure the patient’s
skin is properly prepped.
If a lead or electrode cannot be replaced,
select another ST lead for processing.
Cannot Derive
STVM
Cannot Derive
STCVM
At least one of the 12 leads is
disconnected.
or
STVM was invalid at the time
the reference value was saved.
Inspect and replace defective cables and
wires.
Reapply gel on reusable electrode(s).
Reapply electrodes. Make sure the patient’s
skin is properly prepped.
STCVM: Do a Save Reference.
MultiMed 12 Pod
Disconnected
The MultiMed 12 pod is not
connected to the monitor during
12-lead monitoring.
Check the cables and the connection; replace
the cables, if necessary.
NOTE: Not supported when TruST is
enabled.
MultiMed Pod
Disconnected
The MultiMed pod is not
connected to the monitor during
3, 5, or 6-lead monitoring.
Check the cables and the connection;
replace the cables, if necessary.
If the MultiMed pod is not in use, press the
Alarm Silence fixed key.

Table of Contents

Other manuals for Dräger Infinity Delta

Related product manuals