SPECIALTY MENUS
VF8 DELTA/DELTA XL/KAPPA 2-13
Quick Reference Table – OR Menu
OR Function Description Settings
The OR menu
OR Activates the OR menu functions.
NOTE: This function is a locked
option. It is installed at the time of
purchase from perioperative care unit
or it can be installed locally by
DrägerService or your hospital’s
technical personnel.
• ON – OR functions are enabled.
• OFF – The monitor reverts to normal
functions; The settings Cardiac
Bypass and NBP Chime are
ghosted.
Cardiac Bypass Configures the monitor for use during
cardiac surgery.
• ON – Suspends all patient-
monitoring alarms (except ventilator
alarms), NBP interval
measurements, and arrhythmia
detection.
• OFF – The monitor reverts to normal
functions.
NBP Chime Controls whether an attention tone
sounds upon completion of an NBP
measurement, see chapter 13 for
more information).
• ON – Attention tone sounds when
NBP measurement is complete.
• OFF – No tone sounds when NBP
measurement is complete.
Alarm Volume Sets the alarm volume. • 10 - 100% in increments of 10
•OFF
Large IBP-Mean
Display
Determines the relative size of the
mean pressure value in invasive
pressure parameter boxes.
• ON – The mean IBP value is larger
than the systolic and diastolic IBP
values.
• OFF – The systolic, diastolic and
mean IBP values are the same size.
Attention Tone
Volume
Sets the attention tone volume • Off, 5, 10 - 100% in increments of 10
HR Source Derives heart rate from various
sources (see page 8-23).
NOTE: This function is useful during
electrosurgery when artifact makes
the ECG channel unavailable.
Click on one of the following settings to
determine the Heart Rate source:
•ECG
• ART
•SpO2
• AUTO
Filter Determines sensitivity to noise,
artifact, and other signal distortion
(see page 8-20).
NOTE: The ESU setting automatically
disables pacer detection.
•OFF
• ESU
•Monitor
ARR Monitoring Determines the number of arrhythmia
events you can monitor
(see page 9-5).
•OFF
• Basic
•Full