8-20 DELTA/DELTA XL/KAPPA VF8
8 ECG
AND HEART RATE
Cable Type When set to Auto Detect, the
monitor detects the number of
connected lead wires via a
MultiMed pod automatically. If
Auto Detect mode does not
detect the connected lead wires
(the ECG waveform is not
displayed or the message ECG
Leads Invalid or V+ Lead OFF
appears in the upper left corner
of your screen), you can
manually select the cable type.
The setting 10 (12 Lead)
denotes a combination of a
6-lead wire set and a 4-chest
lead wire set, or a 5-lead wire
set and 5-chest lead wire set for
12-lead monitoring.
NOTE: Verify the Cable Type
setting whenever you begin
ECG monitoring (the monitor
remembers the previous Cable
Type setting).
NOTE: The TruST selection
becomes unavailable when you
connect a MultiMed 6, MultiMed
Plus or a MultiMed Plus OR
cable and the Cable Type
setting is set to either: 3, or 5.
NOTE: The selection 10
(12 lead) is only available when
a MultiMed 12 pod is
connected.
Auto Detect (default) – the monitor
compensates automatically for one
disconnected neutral lead. (This feature is
available only with the Auto Detect setting.)
•3
•5
•6
10 (12 Lead)
NOTE: The selected cable type determines how
many ECG leads are processed regardless of
which ECG cable is connected. For example, if
you connect a MultiMed 6, MultiMed Plus or
MultiMed Plus OR with a 6 lead wire set, but the
Cable Type is set to ‘5’, the monitor will only
process and display 7 leads. The V+ lead will
not be displayed, trended or broadcast to the
network in this case. However, the V+ lead will
still be displayed in the Show All screen. To
display all 8 leads you must set the Cable Type
to ‘6’.
ECG processing for QRS detection and
arrhythmia are still controlled with the QRS/ARR
Select menu setting (see page 8-23) and the
Lead Setup menu selections for Channel 1 and
Channel 2.
The ECG Options Submenu
This submenu allows you to configure the following functions.
Filter Controls the channel bandwidth
and displays a banner in the
ECG1 channel if the selection is
set to OFF or ESU.
No banner is displayed if you
select Monitor.
CAUTION: When the ESU
block or the MultiMed Plus OR
cable is in use, the detection of
pacemaker spikes could be
compromised.
Click on OFF for maximum bandwidth and
greatest sensitivity to noise or artifact.
Click on Monitor (default) to reduce baseline
drift, muscle artifact, and power line
interference (recommended for standard
monitoring, display, recording, and analog
output).
Click on ESU to reduce signal distortion from
electrosurgical units (see page 8-6 for
information on electrosurgical safety).
Menu
Selection
Description Available Settings