ECG SETUP MENU
VF8 DELTA/DELTA XL/KAPPA 8-21
Pacer
Detection
NOTE: See
page page 8-4
and page 2-21
for more
information on
pacemaker
detection.
Determines the monitor’s ability
to identify pacemaker pulse.
Allows user to enable/disable
pacer detection or choose more
advanced Fusion selection.
Caution: Fusion mode pacer
detection is not intended for use
with large-signal, unipolar
pacemakers. It is intended for
use only with biphasic
pacemakers. Please observe
the following:
• Select Fusion mode only in
situations where it becomes
necessary to suppress
repeated false asystole and/or
false low heart rate alarms.
• Before selecting Fusion mode,
be certain that the patient has
a biphasic pacemaker
(external or implanted) and
that it is accurately
programmed as appropriate
for that patient.
• Do not use Fusion mode if you
are uncertain as to what type
of pacemaker is being used.
In basic mode: (see page 2-21)
•Select ON to enable pacemaker detection.
•Select OFF (default) to disable pacemaker
detection.
In advanced mode: (see page 2-21)
•Select ON to enable pacemaker detection.
•Select OFF (default) to disable pacemaker
detection.
•Select Fusion to enable pacer detection,
but minimize pacer tail rejection to reduce
missed detection of pseudo-fused paced
beats, which results in false asystole alarms.
QRS Sync
Marker
The vertical white line that is
displayed on each detected
QRS complex indicates when it
is safe to perform synchronized
cardioversion.
NOTE: Sync output pulses can
trigger the timing of defibrillators
during synchronized
cardioversion.
• Connect the device to the output marked Sync
on the right side of the monitor.
•Select ON to enable the QRS Sync marker.
•Select OFF to disable the QRS Sync marker.
Pulse Tone
Source
Selects the ECG or SpO2 signal
as the source for the pulse tone
A blinking heart () displays in
the parameter box.
• Click on ECG to use the ECG signal as the
source for the pulse tone.
• Click on SpO2 to use the SpO
2
signal as the
source for the pulse tone.
Pulse Tone
Volume
Regulates pulse tone volume • Click on OFF to silence the pulse tone.
• Click on volume (5 to 100 %) to regulate the
pulse tone.
TruST 12-
Lead
Allows the monitor to perform
12-lead monitoring when a
MultiMed 6 pod is used.
•Select ON to enable TruST 12-lead monitoring.
•Select OFF (default) to disable TruST 12-lead
monitoring.
Menu
Selection
Description Available Settings