EasyManua.ls Logo

Fluke 990DSL - Leakage Stress Test; Loop Devices Test Results; Setting up a Leakage Stress Test

Fluke 990DSL
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
POTS Testing and Fault Location
The POTS Toolbox
4
4-37
Loop Devices Test Results
To run the Loop Devices Test, select it on the POTS Toolbox and press K.
Upon completion of the test, the tester displays a result screen that lists each
loop device and indicates whether or not it was detected on the test pair. A
Y
indicates that the device was detected, and an
N indicates that the device was
not detected.
Leakage Stress Test
The Leakage Stress test is a continuous resistance test that uses a higher test
voltage than normal, in order to “punch through” a potential metallic wire fault
that has oxidized over time. During normal resistance testing, this type of fault
typically does not show up on a spare pair. However, it does gradually appear
under sustained application of C.O. battery voltage after being cut into service
as a supposedly “good” pair.
Setting Up a Leakage Stress Test
To set up a Leakage Stress test, select the test on the POTS Toolbox and press
D(
Setups). The test has one setup parameter, which is listed in Table 4-6.
Table 4-6. Leakage Stress Test Setup Parameter
Parameter Settings (default in bold)
Leakage Fault Pass Threshold 500 ke

Table of Contents

Related product manuals