EasyManua.ls Logo

Fluke 990DSL - Pair 1 - 2 Difference Test

Fluke 990DSL
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TDR Testing and Fault Location
The TDR Toolbox
6
6-13
acy73s.eps
Figure 6-3. Pair 1 and 2 Compare Test Result Screen
Pair 1is always depicted as the lower trace and shown as a solid line. Pair 2 is
the upper trace, which is shown as a dotted line. In Figure 6-3, note that the
trace for Pair 2 has a minor reflection (it is marked by the reflection cursor)
that indicates a gauge change at a splice point.
Pair 1 - 2 Difference Test
This test is similar to the Pair 1 and 2 Compare test. The main difference
between the two tests is that in the Pair 1-2 Difference test, the CopperPro
internally computes the mathematical difference between the two waveforms
and displays the difference as a single trace. This presentation is especially
helpful for helping you see minor differences between two pairs—one good
and one suspect.
Pair 1 waveform data is used as the reference, and Pair 2 data is subtracted
from it to obtain the “difference” waveform. If both pairs are identical (have
exactly the same waveforms), the result is a straight line with no reflections.
The TDR softkeys operate in the same manner as those on the Pair 1 test result
screen (see “Pair 1 Test” and Table 6-3 for descriptions).

Table of Contents

Related product manuals