System Overview
2-79
Other status indicators
AVR: The Automatic Voltage Regulation (AVR) feature will compensate for excessively low input
voltages, without using battery power.When the AVR symbol is illuminated on the LCD, the unit is in
Boost mode, using the AVR feature.
UPS configuration
Unit sensitivity
Adjust the sensitivity of the UPS to control when it will switch to battery power; the higher the sensitivity, the more often the
unit will switch to battery power.
1. Ensure the unit is connected to utility power, but is OFF.
2. Press and hold the POWER button for six seconds. The LOAD CAPACITY bar will flash on and off, indicating that the unit is in
programming mode.
3. Press POWER again to rotate through the menu options. Stop at selected sensitivity. The unit will beep to confirm the
selection.
The Medium sensitivity is the model used in Brivo OEC 715/785/865