Installation
 
                   3-41   
The following procedure produces X-rays. Use the appropriate personal protective 
equipment to protect yourself from X-ray exposure. 
 
Image Orientation 
Using the line pair tool as a phantom, take exposure in normal Fluoro mode and verify the following 
functions: 
•  Image Rotation key  rotates the image clockwise and counter-clockwise. 
•  Image Reverse key  inverts and reinverts the image. 
 
Field size  
Place something on the Image Intensifier. Press field size key   repeatedly, verifying ability to 
select any of the three available sizes: NORM, MAG1 and MAG2. 
Collimator 
Press each of the following collimator switches while taking exposure and verify each of the 
following collimator actions. 
•  Shutter Collimator rotate.   
•  Iris Collimator opens and closes.   
•  Shutter Collimator open and close.   
 
kVp , mA/mAs  
Press the kVp and mA/mAs keys and verify that the technique can be adjusted manually. And 
 LED will illuminate indicating Auto.