Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-42
Pulse Fluoroscopy
Select Pulsed normal Fluoro mode via the C-Arm control panel. The X-Ray Exposure indicator keeps
lighting when taking an exposure.
If the pulse mode is enabled, the amount of accumulated time will be the actual exposure
time, which depends on the time of pressing X-ray switch, length of pulses and the
number of pulses per second.
1pulse count to
1pps : 50 milliseconds / 1 pulse approximately
2pps : 50 milliseconds / 1 pulse approximately
4pps : 50 milliseconds / 1 pulse approximately
8pps : 50 milliseconds / 1 pulse approximately
Fluoro time, kVp, mA/mAs Display
Clearly Display Fluoro time, kVp and mA/mAs values.
Alarm Reset
Press and hold Alarm Reset key for approximately 2 seconds. Verify Fluoro time displayed on
the C-Arm control panel resets to zero.
X-ray Emission Indicator
In the working mode, when the footswitch/handswitch is activated to take an exposure, the
indictor will light.
X-Ray off Indicator
Key switch is placed in X-ray off position ( ) or standby position ( ), X-ray is disabled and
this indictor will light.
Radiographic Mode
Select Radiography mode key and verify that the indicator lights.
Low Dose
Select the Low Dose key and verify that the indicator lights.