Calibration
4-25
e. Loosen 3 camera secure screws.
2. Log in software service mode, in calibration TAB, click “Beam alignment…”
3. Use the 4 Camera alignment screws to adjust the camera center. Make sure to unscrew one screw
first, and then fasten the screw on the other side to center camera. Perform camera centering
check and repeat adjusting camera alignment screws as many times as necessary.
4. When the camera centering verification meets the specification, fasten the alignment screws and
camera secure screws.
5. After fasten secure screws, recheck the alignment result. If fail, redo calibration.
Securing Screw
Camera Alignment Screws (there
are 4 screws with 90 degrees apart)