Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-8
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0104001
Warning
DC is Error
Generator power exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. Check input voltage and
power supply.
2. Check DC Bus fuse and
related cables.
3. Check the IGBT.
4. Check the KV control
board.
5. Check the monoblock.
0104002
Warning
IGBT is error
Generator power exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. Check drive wires
between the kV control
board and IGBT board.
2. Check the IGBT.
3. Check the KV control
board and power supply to
KVCB.
0104003
Warning
KV is high than
120KV
Generator exception. Exposure
is disabled. Please restart the
system. If this issue persists,
please call for service.
KV need be calibrated
0104004
Warning
KV is lower than
35kv
Generator exception. Exposure
is disabled. Please restart the
system. If this issue persists,
please call for service.
1. Check input voltage and
power supply.
2. Check DC Bus fuse and
related cables.
3. Check the IGBT.
4. Check the KV control
board.
5. Redo KV calibration.
6. Check the monoblock.
0104005
Warning
KV is unbalance
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check the bundle cable.
2. Check KV_feedback
signal on kV control board.
3. Check the monoblock.
0105001
Warning
Filament is over
current by HW
monitor
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check DC power supplier
and fuse F6 on the Mother
board (backplane).
2. Check related cables.
3. Check the MCB.
4. Check the FDB.
5. Check the monoblock.
0105002
Warning
Inverter is over
current
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1.Re-do Calibration.
2. Check DC power supplier
and fuse F5 on the Mother
board (backplane).
2. Check the FDB.
3. Replace the monoblock.